Computational design and evaluation of a quad-MOSFET device for quality control of therapeutic accelerator-based neutron beams

Publication Name

Radiation Measurements

Abstract

Accurate real-time monitoring of neutron beams and distinguishing between thermal, epithermal and fast neutron components in the presence of a photon background is crucial for the effectiveness of accelerator-based boron neutron capture therapy (AB-BNCT). In this work, we propose an innovative quadruple metal–oxide–semiconductor field-effect transistor (MOSFET) device for real-time, cost-effective beam quality control; one detector is kept uncovered while the other three are covered with either a B4C, cadmium and B4C or polyethylene converter. Individual MOSFET converter configurations were optimised via Monte Carlo simulations to maximise signal selectivity across neutron energy spectra. Results demonstrate the quad-MOSFET device's efficacy in quantifying changes in neutron flux, underscoring its potential as a useful instrument in the AB-BNCT quality control process.

Open Access Status

This publication may be available as open access

Volume

177

Article Number

107253

Funding Sponsor

Australian Institute of Nuclear Science and Engineering

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1016/j.radmeas.2024.107253