Comparison of SOI microdosimeter and tissue equivalent proportional counter measurements at the CERF facility

RIS ID

69539

Publication Details

Prokopovich, D. A., Reinhard, M. I., Taylor, G. C., Hands, A. & Rosenfeld, A. B. (2012). Comparison of SOI microdosimeter and tissue equivalent proportional counter measurements at the CERF facility. IEEE Transactions on Nuclear Science, 59 (Issue 5 Part 3), 2501-2505.

Abstract

The CERN-EU High Energy Reference Field (CERF) facility is used in the calibration of neutron dosimeters for radiation protection applications in aviation and high energy physics. A comparison of the facility’s microdosimetric spectra obtained with a new Silicon on Insulator (SOI) Microdosimeter and a HAWK Tissue Equivalent Proportional Counter (TEPC) is presented. Experimental data obtained with both devices indicates the presence of a small charged particle flux within the neutron dominated field which makes a non negligible contribution to the dose equivalent.

Please refer to publisher version or contact your library.