Measurement of scatter radiation from the GE Infinia Hawkeye 4 SPECT/CT system

RIS ID

34095

Publication Details

Harvey, S. B. (2010). Measurement of scatter radiation from the GE Infinia Hawkeye 4 SPECT/CT system. Australasian physical and engineering sciences in medicine / supported by the Australasian College of Physical Scientists in Medicine and the Australasian Association of Physical Sciences in Medicine, 33 (2), 193-197.

Abstract

In this study the cumulative scatter dose from a typical monthly SPECT/CT workload on a new GE Infinia Hawkeye 4 system was measured by means of a dosimeter badge survey. Doses were found to be well within regulatory limits. Scatter dose rates attributable solely to the CT component were also measured experimentally for comparison with the manufacturer’s isocontour plot. The measured CT dose rate isocontours were found to be of similar magnitude and location to those quoted by the manufacturer. Data generated from this study will form a valuable baseline for the monitoring of equipment performance and for future scatter radiation measurements.

Please refer to publisher version or contact your library.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1007/s13246-010-0013-9