Piezoresistive sensors for force mapping of hip-prostheses

RIS ID

99127

Publication Details

Correia, V., Sencadas, V., Martins, M. S., Ribeiro, C., Alpuim, P., Rocha, J. G., Morales, I., Atienza, C. & Lanceros-Méndez, S. (2013). Piezoresistive sensors for force mapping of hip-prostheses. Sensors and Actuators A: Physical, 195 133-138.

Abstract

The success of artificial prosthetic replacements depends on the fixation of the artificial prosthetic component after being implanted in the thighbone. This work shows a smart prosthesis based on highly sensitive silicon thin-film piezoresistive sensors attached to a hip prosthesis. The performance of the sensors for this application is studied and compared to commercial strain gauge sensors. Mechanical stress-strain experiments were performed in compressive mode, during 10,000 cycles and data was acquired at mechanical vibration frequencies of 0.5 Hz, 1 Hz and 5 Hz, and sent to a computer by means of a wireless link. The results show that there is a decrease in sensitivity of the thin-film silicon piezoresistive (n-type nanocrystalline Si) sensors when they are attached to the prosthesis, however this decrease does not compromise its monitoring performance. The sensitivity, compared to that of commercial strain gauges, is much larger due to their higher gauge factor (-23.5), when compared to the gage factor of commercial sensors (2).

Please refer to publisher version or contact your library.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1016/j.sna.2013.03.013