Incorporating clinical imaging into the delivery of microbeam radiation therapy
Publication Name
Applied Sciences (Switzerland)
Abstract
Synchrotron microbeam radiation therapy is a promising pre-clinical radiation treatment modality; however, it comes with many technical challenges. This study describes the image guidance protocol used for Australia’s first long-term pre-clinical MRT treatment of rats bearing 9L gliosarcoma tumours. The protocol utilises existing infrastructure available at the Australian Synchrotron and the adjoining Monash Biomedical Imaging facility. The protocol is designed and optimised to treat small animals utilising high-resolution clinical CT for patient specific tumour identification, coupled with conventional radiography, using the recently developed SYNCMRT program for image guidance. Dosimetry performed in small animal phantoms shows patient dose is comparable to standard clinical doses, with a CT associated dose of less than 1.39 cGy and a planar radiograh dose of less than 0.03 cGy. Experimental validation of alignment accuracy with radiographic film demonstrates end to end accuracy of less than ±0.34 mm in anatomical phantoms. Histological analysis of tumour-bearing rats treated with microbeam radiation therapy verifies that tumours are targeted well within applied treatment margins. To date, this technique has been used to treat 35 tumour-bearing rats.
Open Access Status
This publication is not available as open access
Volume
11
Issue
19
Article Number
9101
Funding Number
14127
Funding Sponsor
Australian Nuclear Science and Technology Organisation