RIS ID

103953

Publication Details

Pogson, E. M., Begg, J., Jameson, M. G., Dempsey, C., Latty, D., Batumalai, V., Lim, A., Kandasamy, K., Metcalfe, P. E. & Holloway, L. C. (2015). A phantom assessment of achievable contouring concordance across multiple treatment planning systems. Radiotherapy and Oncology, 117 (3), 438-441.

Abstract

In this paper, the highest level of inter- and intra-observer conformity achievable with different treatment planning systems (TPSs), contouring tools, shapes, and sites have been established for metrics including the Dice similarity coefficient (DICE) and Hausdorff Distance. High conformity values, e.g. DICEBreast_Shape = 0.99 ± 0.01, were achieved. Decreasing image resolution decreased contouring conformity.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1016/j.radonc.2015.09.022