Preston, R. M., Eberhardt, J. & Tickner, J. R. (2013). Neutron generator burst timing measured using a pulse shape discrimination plastic scintillator with silicon photomultiplier readout. Journal of Instrumentation, 8 P12005-1-P125001-12.

Abstract

An EJ-299-34 plastic scintillator with silicon photomultiplier (SiPM) readout was used to measure the fast neutron output of a pulsed Thermo-Fisher A-325 Deuterium-Tritium sealed tube neutron generator (STNG). The SiPM signals were handled by a prototype digital pulse processing system, based on a free-running analogue to digital converter feeding a digital signal processor (DSP). Pulse shape discrimination was used to distinguish between detected fast-neutrons and gammas. Pulse detection, timing, energy and shape were all processed by the DSP in real-time.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1088/1748-0221/8/12/P12005