posted on 2024-11-15, 22:18 authored by P D Bradley, Anatoly RozenfeldAnatoly Rozenfeld
Reverse-biased silicon p-n junction arrays have been proposed as microdosimetry detectors. The tissue equivalence of such detectors in boron neutron capture therapy (BNCT) is discussed. A comparison of the range-energy relationships of H, He, C, and Li ions in tissue (ICRU-muscle) and silicon is given. A simple geometrical scaling (~0.63) of linear dimensions is required to convert microdosimetric energy deposition measurements performed in silicon to equivalent deposition in tissue. The Monte Carlo technique is used to examine energy deposition for two simple geometrical cases applicable to BNCT.

History

Citation

This article was originally published as Bradley, PD and Rosenfeld, AB, Tissue equivalence correction for silicon microdosimetry detectors in boron neutron capture therapy, Medical Physics, 25(11), November 1998, 2220. Copyright American Association of Physicists in Medicine. Original journal available here.