Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI-Linac prototypes

journal contribution
posted on 2024-11-16, 04:25 authored by Sarah Alnaghy, Jarrad Begg, Trent Causer, Thahabah Alharthi, Laura Glaubes, Bin Dong, Armia George, Lois HollowayLois Holloway, Peter MetcalfePeter Metcalfe
Purpose Longitudinal magnetic fields narrow beam penumbra and tighten lateral spread of secondary electrons in air cavities, including lung tissue. Gafchromic® EBT3 film was used to investigate differences between penumbra in solid water and solid lung, without a magnetic field (0 T) and with two field strengths (0.9 and 1.5 T). Methods The first prototype of the Australian MRI‐linac consisted of a 1.5 T Siemens Sonata MRI and Varian industrial linatron (nominal 4 MV). The second prototype replaced the Sonata with a 1.0 T Agilent split‐bore magnet. Measurements were completed at 0.9 T to maintain the same source‐to‐surface distance between set‐ups. Gammex‐rmi® solid water with 50 mm of CIRS solid lung inserted as a lung cavity was positioned inside each magnet. This was compared to the same set‐up with solid water only, where film measurements were completed at solid water equivalent depths corresponding to entrance interface/mid/exit interface positions of solid lung from the first set‐up. Multileaf collimator (MLC)‐defined field sizes were set to 3 × 3 cm2 and 10 × 10 cm2. The 80%–20% penumbral width was determined. Results Under 1.5 T conditions, penumbra narrowing occurred up to 4.4 ± 0.1 mm compared to 0 T. As expected, the effect was less for 0.9 T, which resulted in a maximum narrowing of 2.5 ± 0.1 mm. Exit profile penumbra were more affected than entrance penumbra by up to 2.6 ± 0.2 mm. The 1.5 T field brought the solid water and lung penumbral widths more into alignment by a maximum difference of 0.4 ± 0.1 mm. Conclusions The trimming of penumbral widths due to magnetic fields in solid water and lung was demonstrated and compared to 0 T. The 0.9 and 1.5 T field trimmed the penumbra by up to 2.5 ± 0.1 mm and 4.4 ± 0.1 mm respectively.

Funding

The Australian MRI-Linac Program: Improving cancer treatment through real-time image guided adaptive radiotherapy

National Health and Medical Research Council

Find out more...

History

Citation

Alnaghy, S. J., Begg, J., Causer, T., Alharthi, T., Glaubes, L., Dong, B., George, A., Holloway, L. & Metcalfe, P. (2018). Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI-Linac prototypes. Medical Physics, 45 (1), 479-487.

Journal title

Medical Physics

Volume

45

Issue

1

Pagination

479-487

Language

English

RIS ID

118667

Usage metrics

    Categories

    Keywords

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC