posted on 2024-11-15, 07:23 authored by G Gambarini, M Carrara, C Tenconi, N Mantaut, M Borroni, Dean CutajarDean Cutajar, Marco PetaseccaMarco Petasecca, I Fuduli, Michael LerchMichael Lerch, E Pignoli, Anatoly RozenfeldAnatoly Rozenfeld
MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Results show that dual-MOSkin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.

History

Citation

Gambarini, G., Carrara, M., Tenconi, C., Mantaut, N., Borroni, M., Cutajar, D., Petasecca, M., Fuduli, I., Lerch, M., Pignoli, E. & Rosenfeld, A. (2014). Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: In phantom feasibility study. Applied Radiation and Isotopes, 83 (Part C), 222-226.