A Geant4 simulation modelling the passive Biological beamline at the Heavy Ion Medical Accelerator Chiba (HIMAC) is described and validated against experimental measurements for a mono-energetic and spread out Bragg peak (SOBP) 12 C beam. Comparisons to experimental data showed good agreement for both lateral and depth dose profiles. It was found that simplifying the simulation model by generating a cone beam instead of using the wobbler magnets provided a good approximation for both lateral and depth dose measurements.
History
Citation
Bolst, D., Tran, L. T., Guatelli, S., Matsufuji, N. & Rosenfeld, A. B. (2019). Modelling the Biological Beamline at HIMAC using Geant4. Journal of Physics: Conference Series, 1154 (Conference 1), 1-6.