Innovative detectors for quality assurance dosimetry in SBRT of stationary and movable targets

Download (790.61 kB)
journal contribution
posted on 2024-11-15, 14:09 authored by Matthew NewallMatthew Newall, Marco PetaseccaMarco Petasecca, Mitchell Duncan, Iolanda Fuduli, Khalsa Al shukaili, Jeremy T Booth, Paul J Keall, Stephanie Corde Tehei, V Pereveratylo, Michael LerchMichael Lerch, Anatoly RozenfeldAnatoly Rozenfeld
The high spatial and temporal resolution 2D monolithic silicon detector arrays M512 and DUO for quality assurance (QA) in real time motion adaptive radiotherapy (ART) have been developed. The DUO array possesses a spatial resolution of 0.2 mm and has demonstrated agreement within 5% with EBT3 film measurements of 6MV linac beam profiles for field sizes 1 × 1 cm2 and SRS cone diameter 0.5 cm. Dynamic characterisation of the M512 for QA in real time ART evaluated the performance of M512 for small fields while the detector experiences periodic motion. It was demonstrated with M512 that MLC tracking with Calypso electromagnetic array compensates for the periodic motion and improves the agreement between static and dynamic beam profiles for field size 1 × 1 cm2 from within 75% in the penumbra to within 11% agreement. The dynamic profile is returned to a similar distribution as the static case.

History

Related Materials

  1. 1.
    ISSN - Is published in 1742-6588

Citation

Newall, M. K., Petasecca, M., Duncan, M., Fuduli, I., Al shukaili, K., Booth, J. T., Keall, P., Corde, S., Pereveratylo, V., Lerch, M. & Rosenfeld, A. B. (2017). Innovative detectors for quality assurance dosimetry in SBRT of stationary and movable targets. Journal of Physics: Conference Series, 777 (1), 012014-1-012014-4.

Language

English

RIS ID

112336

Journal title

Journal of Physics: Conference Series

Volume

777

Issue

1

Usage metrics

    Categories

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC