Improvements in dose calculation accuracy for small off-axis targets in high dose per fraction tomotherapy

Download (6.09 MB)
journal contribution
posted on 2024-11-15, 05:46 authored by Nicholas HardcastleNicholas Hardcastle, Adam Bayliss, Jeannie Wong, Anatoly RozenfeldAnatoly Rozenfeld, Wolfgang Tome
Purpose: A recent field safety notice from TomoTherapy detailed the underdosing of small, off-axis targets when receiving high doses per fraction. This is due to angular undersampling in the dose calculation gantry angles. This study evaluates a correction method to reduce the underdosing, to be implemented in the current version (v4.1) of the TomoTherapy treatment planning software. Methods: The correction method, termed "Super Sampling" involved the tripling of the number of gantry angles from which the dose is calculated during optimization and dose calculation. Radiochromic film was used to measure the dose to small targets at various off-axis distances receiving a minimum of 21 Gy in one fraction. Measurements were also performed for single small targets at the center of the Lucy phantom, using radiochromic film and the dose magnifying glass (DMG). Results:Without super sampling, the peak dose deficit increased from 0% to 18% for a 10 mm target and 0% to 30% for a 5 mm target as off-axis target distances increased from 0 to 16.5 cm. When super sampling was turned on, the dose deficit trend was removed and all peak doses were within 5% of the planned dose. For measurements in the Lucy phantom at 9.7 cm off-axis, the positional and dose magnitude accuracy using super sampling was verified using radiochromic film and the DMG. Conclusions: A correction method implemented in the TomoTherapy treatment planning system which triples the angular sampling of the gantry angles used during optimization and dose calculation removes the underdosing for targets as small as 5 mm diameter, up to 16.5 cm off-axis receiving up to 21 Gy.

History

Citation

Hardcastle, N., Bayliss, A., Wong, J., Rosenfeld, A. B. & Tome, W. A. (2012). Improvements in dose calculation accuracy for small off-axis targets in high dose per fraction tomotherapy. Medical Physics, 39 (8), 4788-4794.

Journal title

Medical Physics

Volume

39

Issue

8

Pagination

4788-4794

Language

English

RIS ID

63493

Usage metrics

    Categories

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC