Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam

Download (119.77 kB)
journal contribution
posted on 2024-11-15, 10:18 authored by G I Kaplan, Anatoly RozenfeldAnatoly Rozenfeld, B J Allen, J T Booth, Martin Carolan, A Holmes-Siedle
Measurement of the lateral profile of the dose distribution across a narrow x-ray microbeam requires a dosimeter with a micron resolution. We investigated the use of a MOSFET dosimeter in an ‘‘edge-on’’ orientation with the gate insulating oxide layer parallel to the direction of the beam. We compared results using this technique to Gafchromic film measurements of a 200 micrometer wide planar x-ray microbeam. The microbeam was obtained by using a vernier micrometerdriven miniature collimator attached to a Therapax DXT300 x-ray machine operated at 100 kVp . The ‘‘edge-on’’ application allows utilization of the ultra thin sensitive volume of the MOSFET detector. Spatial resolution of both the MOSFET and Gafchromic film dosimeters appeared to be of about 1 micrometer. The MOSFET dosimeter appeared to provide more uniform dose profiles with the advantage of on-line measurements.