University of Wollongong
Browse

Effective area and charge density of iridium oxide neural electrodes

Download (650.1 kB)
journal contribution
posted on 2024-11-16, 05:59 authored by Alexander Harris, Antonio G Paolini, Gordon WallaceGordon Wallace
The effective electrode area and charge density of iridium metal and anodically activated iridium has been measured by optical and electrochemical techniques. The degree of electrode activation could be assessed by changes in electrode colour. The reduction charge, activation charge, number of activation pulses and charge density were all strongly correlated. Activated iridium showed slow electron transfer kinetics for reduction of a dissolved redox species. At fast voltammetric scan rates the linear diffusion electroactive area was unaffected by iridium activation. At slow voltammetric scan rates, the steady state diffusion electroactive area was reduced by iridium activation. The steady state current was consistent with a ring electrode geometry, with lateral resistance reducing the electrode area. Slow electron transfer on activated iridium would require a larger overpotential to reduce or oxidise dissolved species in tissue, limiting the electrodes charge capacity but also reducing the likelihood of generating toxic species in vivo.

Funding

ARC Centre of Excellence - Australian Centre for Electromaterials Science

Australian Research Council

Find out more...

ARC Centre of Excellence for Electromaterials Science

Australian Research Council

Find out more...

History

Citation

Harris, A. R., Paolini, A. G. & Wallace, G. G. (2017). Effective area and charge density of iridium oxide neural electrodes. Electrochimica Acta, 230 285-292.

Journal title

Electrochimica Acta

Volume

230

Pagination

285-292

Language

English

RIS ID

112477

Usage metrics

    Categories

    Keywords

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC