University of Wollongong
Browse

Determining the longitudinal accuracy and reproducibility of T1 and T2 in a 3T MRI scanner

journal contribution
posted on 2024-11-17, 12:42 authored by Madeline E Carr, Kathryn E Keenan, Robba Rai, Peter Metcalfe, Amy Walker, Lois Holloway
Purpose: To determine baseline accuracy and reproducibility of T1 and T2 relaxation times over 12 months on a dedicated radiotherapy MRI scanner. Methods: An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T1 was measured using inversion recovery (T1-IR) and variable flip angle (T1-VFA) sequences and T2 was measured using a multi-echo spin echo (T2-SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T1 and T2 times compared to reference values. Reproducibility was measured by the coefficient of variation (CV) of T1 and T2 measurements across monthly scans. Accuracy and reproducibility were mainly assessed on vials with relaxation times expected to be in physiological ranges at 3T. Results: A strong linear correlation between measured and reference relaxation times was found for all sequences tested (R2> 0.997). Baseline bias (and CV[%]) for T1-IR, T1-VFA and T2-SE sequences were +2.0% (2.1), +6.5% (4.2), and +8.5% (1.9), respectively. Conclusions: The accuracy and reproducibility of T1 and T2 on the scanner were considered sufficient for the sequences tested. No longitudinal trends of variation were deduced, suggesting less frequent measurements are required following the establishment of baselines.

History

Journal title

Journal of Applied Clinical Medical Physics

Volume

22

Issue

11

Pagination

143-150

Language

English

Usage metrics

    Categories

    No categories selected

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC