University of Wollongong
Browse

Clinical validation of an in-house EPID dosimetry system for IMRT QA at the Prince of Wales Hospital

Download (453.98 kB)
journal contribution
posted on 2024-11-15, 14:16 authored by M Tyler, P Vial, Peter MetcalfePeter Metcalfe, S Downes
In this study a simple method using standard flood-field corrected Electronic Portal Imaging Device (EPID) images for routine Intensity Modulated Radiation Therapy (IMRT) Quality Assurance (QA) was investigated. The EPID QA system was designed and tested on a Siemens Oncor Impression linear accelerator with an OptiVue 1000ST EPID panel (Siemens Medical Solutions USA, Inc, USA) and an Elekta Axesse linear accelerator with an iViewGT EPID (Elekta AB, Sweden) for 6 and 10 MV IMRT fields with Step-and-Shoot and dynamic-MLC delivery. Two different planning systems were used for patient IMRT field generation for comparison with the measured EPID fluences. All measured IMRT plans had >95% agreement to the planning fluences (using 3 cGy / 3 mm Gamma Criteria) and were comparable to the pass-rates calculated using a 2-D diode array dosimeter.

History

Citation

Tyler, M., Vial, P., Metcalfe, P. & Downes, S. (2013). Clinical validation of an in-house EPID dosimetry system for IMRT QA at the Prince of Wales Hospital. Journal of Physics, 444 (012043), 1-4.

Journal title

Journal of Physics: Conference Series

Volume

444

Issue

1

Language

English

RIS ID

82328

Usage metrics

    Categories

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC