Purpose: This paper presents initial experimental results from a prototype of high dose rate (HDR) BrachyView, a novel in-body source tracking system for HDR brachytherapy based on a multipinhole tungsten collimator and a high resolution pixellated silicon detector array. The probe and its associated position estimation algorithms are validated and a comprehensive evaluation of the accuracy of its position estimation capabilities is presented.
Funding
TEMPERATURE STUDY OF CUTTING TOOLS IN CONVENTIONAL CUTTING SPEED RANGE AND IN HIGH-SPEED MACHINING CONDITIONS.
Coordenação de Aperfeicoamento de Pessoal de Nível Superior
Safavi-Naeini, M., Han, Z., Alnaghy, S., Cutajar, D., Petasecca, M., Lerch, M. L. F., Franklin, D. R., Bucci, J., Carrara, M., Zaider, M. & Rosenfeld, A. B. (2015). BrachyView, a novel in-body imaging system for HDR prostate brachytherapy: Experimental evaluation. Medical Physics, 42 (12), 7098-7107.