posted on 2024-11-15, 15:52 authored by Taylah Brennen, Dean CutajarDean Cutajar, Saree AlnaghySaree Alnaghy, Joseph A Bucci, Andrej Bece, Marco Favoino, M Carriero, Michael LerchMichael Lerch, Anatoly RozenfeldAnatoly Rozenfeld, M Petasseca
BrachyView is a novel in-body imaging system developed with the objective to provide real-time intraoperative dosimetry for low dose rate (LDR) prostate brachytherapy treatments. Seed positions can be reconstructed after implantation by the means of a high-resolution pinhole gamma camera. The obtained dataset is then combined with conventional trans-rectal ultrasound (TRUS) imaging to locate the effective source position within the prostate volume. This study presents a comparison of two data sets containing 96 and 98 I-25 brachytherapy sources and readout by two different pin-hole collimators with 800 m and 500 m, respectively. This study aims to assess the effect on reconstruction of large pinholes which allow for large Field of View (FoV) and statistics of counts on the seed's projections. It was found that 100% of seeds were reconstructed by the means of the 500 m collimator. Results from the 800 m data showed that only 70% of seeds were reconstructed with a total 3D discrepancy of 8.20 mm in respect to CT data and an increased reconstructed position inaccuracy.

History

Citation

Brennen, T., Cutajar, D. L., Alnaghy, S., Bucci, J., Bece, A., Favoino, M., Carriero, M., Lerch, M., Rosenfeld, A. B. & Petasseca, M. (2019). BrachyView: Verification of LDR patient plans - Hardware optimisation. Journal of Physics: Conference Series, 1154 (Conference 1), 012005-1-012005-5.

Journal title

Journal of Physics: Conference Series

Volume

1154

Issue

1

Language

English

RIS ID

135135

Usage metrics

    Categories

    Keywords

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC