Sensitivity variation of doped Fricke gel irradiated with monochromatic synchrotron X-rays between 33.5 and 80 keV

RIS ID

76644

Publication Details

corde, s., Adam, J., Biston, M., Joubert, A., Charvet, A., Esteve, F., Le Bas, J., Elleaume, H. & Balosso, J. (2005). Sensitivity variation of doped Fricke gel irradiated with monochromatic synchrotron X-rays between 33.5 and 80 keV. Radiation Protection Dosimetry, 117 (4), 425-431.

Abstract

An experimental binary radiotherapy proposes the concomitant use of a high-Z compound and synchrotron X rays for enhancing radiation dose selectively in tumours by a photoelectric effect. This study aimed at measuring the resulting dose enhancement in irradiated material. A doped Fricke gel dosemeter model was manufactured with 10 mg ml1 of iodine (Telebrix) or barium (Micropaque). Samples were irradiated with a monochromatic synchrotron beam at 33.5, 50, 65 and 80 keV. The ensuing enhancement of the sensitivity of the dosemeter was derived from the nuclear magnetic resonance relaxation rates measured at different X-ray doses. Our results demonstrate (1) the preservation of a linear relationship between relaxation rates and X-ray doses for dosemeters doped with high-Z atoms and (2) a clear energy-dependent sensitivity enhancement for barium-doped Fricke gels. This enhancement was neither reproducible with iodinated compounds nor clearly related to the expected dose enhancement factor. However 1% barium sulphate in the gel could significantly improve the gel's response when it was irradiated by low-energy X rays.

Please refer to publisher version or contact your library.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1093/rpd/nci313