Buckley, J., Wilkinson, D., Malaroda, A. & Metcalfe, P. (2016). Evaluation of dose from kV cone-beam computed tomography during radiotherapy: A comparison of methodologies. Journal of Physics: Conference Series, 777 (1), 012003-1-012003-4.

Abstract

Three alternative methodologies to the Computed-Tomography Dose Index for the evaluation of Cone-Beam Computed Tomography dose are compared, the Cone-Beam Dose Index, IAEA Human Health Report No. 5 recommended methodology and the AAPM Task Group 111 recommended methodology. The protocols were evaluated for Pelvis and Thorax scan modes on Varian® On-Board Imager and Truebeam kV XI imaging systems. The weighted planar average dose was highest for the AAPM methodology across all scans, with the CBDI being the second highest overall. A 17.96% and 1.14% decrease from the TG-111 protocol to the IAEA and CBDI protocols for the Pelvis mode and 18.15% and 13.10% decrease for the Thorax mode were observed for the XI system. For the OBI system, the variation was 16.46% and 7.14% for Pelvis mode and 15.93% to the CBDI protocol in Thorax mode respectively.

Share

COinS
 

Link to publisher version (DOI)

http://dx.doi.org/10.1088/1742-6596/777/1/012003